The Hologic Fluoroscan InSight 2 is a compact mini C-arm designed specifically for orthopedic and extremity imaging. It delivers high-quality digital fluoroscopy and snapshot images at low dose, making it ideal for the OR, ED, physician offices, ambulatory surgery centers, and sports medicine facilities. With a standard 6"/4" high-sensitivity image intensifier, laser alignment, and a 1k × 1k image matrix, the InSight 2 offers sharp, detailed imaging while keeping workflow simple and efficient. Solid-state storage, a Windows-based interface, and extensive connectivity options support modern digital imaging environments.
Key Features
- 6"/4" high-sensitivity image intensifier
- Laser alignment light for precise positioning
- 1k × 1k image matrix size
- Solid-state hard drive with up to 10,000 images (500 per case)
- Windows 7 Professional–based system
- Multiple imaging modes: snapshot, continuous fluoro, cine
Technical Specifications
| Base System Dimensions | |
|---|---|
| Footprint | 28 × 32 in (71.1 × 81.3 cm) |
| Door Clearance | 30 in (76.2 cm) minimum |
| C-Arm Geometry | |
| Depth | 16 in (40.6 cm) |
| Free Space | 14 in (35.5 cm) |
| Orbital Rotation | 120° |
| Horizontal Travel | 23° |
| Vertical Travel | 23 in (58.4 cm) |
| Pivot Rotation | 380° |
| Panning Motion | 320° |
| X-ray System | |
| X-ray Source | Grounded tungsten anode tube, 7.5 W HF |
| Focal Spot | 0.045 mm |
| Tube kVp Range | 40–75 kVp |
| Tube Current Range | 0.020–0.100 mA |
| Maximum Output | 0.1 mA @ 75 kVp |
| Laser Alignment | 60-second timed or full-time on; IEC Class 1C X-ray Control Modes |
| Detail | Auto IQ (high-resolution mode) |
| Detail | Auto Standard (general imaging) |
| Detail | Manual override mode |
| Image Acquisition & Storage | |
| Standard Image Rate | 30 fps |
| Cine Recording | 25 fps |
| Storage | SSD with 10,000 total images / up to 500 per case |
| Imaging Modes | |
| Detail | Snapshot (single-shot) |
| Detail | Continuous fluoroscopy |
| Detail | Cine digital fluoroscopy |
| Noise Suppression | |
| Selectable levels | Ultra / Auto / High / Med / Low / Off Imaging Features |
| Detail | Automated image processing |
| Detail | Real-time noise reduction, edge enhancement, and auto dose control |
| Detail | Magnify / Zoom / Pan |
| Detail | Brightness and contrast control |
| Detail | Physician preference presets |
| Software & Connectivity | |
| OS | Windows 7 Professional |
| Connectivity | Ethernet wired; optional wireless |
| DICOM | Store (with annotation), Print, Modality Worklist, Storage Commitment |
| Removable Storage | USB and CD/DVD-RAM |
| Export Formats | JPEG, TIFF, DICOM, Fluoroscan/native |
| Dose/Exam Reporting | |
| Structured dose report | DAP (cGy·cm²), Air Kerma (mGy) |
| Detail | Total exposure time (min:sec), longest single exposure, total images acquired and saved |
| Power Requirements | |
| Input Voltage | 100 / 120 / 220 / 240 VAC selectable |
| Frequency | 50/60 Hz |
| Power | up to 750 W |
| Power Failure Protection | Images software-protected |