GE AMX 4

The GE AMX 4 is a classic, battery-powered portable X-ray system designed for dependable radiographic imaging wherever patients need care.

Single Product Request Form

The GE AMX 4 is a classic, battery-powered portable X-ray system designed for dependable radiographic imaging wherever patients need care. Operating entirely from an onboard rechargeable battery pack, the AMX 4 delivers full-power exposures without requiring a wall outlet, making it ideal for ICUs, CCUs, emergency rooms, ORs, orthopedic suites, clinics, pediatric and neonatal units. This system features the MX-75, 275 kHU X-ray tube with a 0.75 mm focal spot, combined with a GE AMX-IV collimator for reliable, high-quality imaging across a wide range of procedures. Variable forward and reverse drive speeds and its fully mobile design allow technologists to position the unit quickly and safely at the bedside. With 24 kVp stations, 30 mAs stations, and consistent 100 mA output, the AMX 4 provides flexible technique selection to meet diverse clinical demands. An LED display and intuitive controls further enhance usability in busy hospital environments.

Key Features

  • Battery-operated portable X-ray system
  • MX-75 275 kHU X-ray tube with 0.75 mm focal spot
  • GE AMX-IV collimator
  • High-quality image performance
  • Forward and reverse variable speed drive
  • Fully mobile, no power connection required for full-power operation
  • 24 kVp stations, 30 mAs stations
  • 100 mA constant output
  • kVp Range: 50–125 kVp
  • mAs Range: 0.4–320 mAs

Technical Specifications

Dimensions
Height76 in (1930 mm)
Width25-3/16 in (640 mm)
Depth45-3/8 in (1153 mm)
Weight1050 lb (477 kg)
BatteryNine 12.9 V batteries connected in series (approx. 116 V at full charge)
Approximate capacity20,000 mAs at 100 kVp under specified conditions
BatteryActual capacity depends on battery condition and use of collimator lamp, electric brakes, rotor, and driving distance
Movements (tube vertical movement measured at focal spot)
Vertical RangeAt least 52.5 in (1334 mm)
Lowest Position26.1 in (663 mm) from floor (max)
Highest Position78.6 in (1996 mm) from floor (min)
Horizontal movement at focal spot relative to column face24 in (610 mm) min to 40 in (1016 mm) max
Tube Column Rotation (from horizontal latch)360°
Tube and Yoke Rotation around Horizontal Arm (from tube port down)360° range; detents at 0°, ±90°, ±180°
Tube Trunnion Rotation (from tube port down)120° range; forward 110°, backward 10°, detents at 0°, 90°
Collimator Rotation (from front with tube port down)180° range; right 90°, left 90°, detents at 0°, 90°
Drive Speeds
Speed 1264 ft (6705 mm) per minute ±25% (horizontal arm secured)
Speed 230–60% of drive speed (maneuvering speed, arm out of latch)
DetailDrive speed measured on smooth, hard, level surface; reduced on inclines, carpet, or soft surfaces
CollimatorMinimum source-to-skin distance >30 cm via skin spacers
CollimatorFull 17 in × 17 in (43 cm × 43 cm) radiographic coverage at 40 in (1.02 m) SID
Minimum inherent filtration2.0 mm Al equivalent at 100 kVp
Light Field IntensityAverage illumination at 100 cm (39.37 in) from focal spot ≥ 16 foot-candles (≈170 lux)