The GE Definium AMX 700 is a battery-powered mobile digital X-ray system designed to deliver high-quality imaging with a flat-panel digital detector and low patient dose. Its compact, highly maneuverable design makes it ideal for imaging in intensive care units, emergency departments, operating rooms, and orthopedic environments where mobility and speed are critical. A flexible positioning arm with 270° column flexibility allows technologists to comfortably position the tube around the patient, minimizing repositioning and improving comfort. The integrated 15" touchscreen display offers intuitive protocol selection and image review, while digital networking capabilities enable rapid image transfer to PACS and enterprise systems. With storage capacity for over 2,000 images and an integrated rotating anode tube, the Definium AMX 700 supports efficient workflow and dependable performance in demanding clinical settings.
Key Features
- 270° column flexibility for versatile positioning
- Easy maneuverability for bedside and tight-space imaging
- 15 in touchscreen display for protocol setup and image control
- Single-piece flat-panel detector
- Instant distribution and access to patient data
- Storage for over 2000 images
- Rotating anode X-ray tube
- Networking connectivity: DICOM 3.0 and IHE compatible
Technical Specifications
| Dimensions | |
|---|---|
| Overall Width | 64.8 cm / 25.5 in (includes front bumper) |
| Width at Midpoint | 64.1 cm / 25.25 in |
| Overall Height | 193 cm / 76 in |
| Tube Height (transport lock over) | 142.2 cm / 56 in |
| Overall Length | 115.6 cm / 45.5 in (includes handle) |
| Base Length | 111.1 cm / 43.75 in |
| Full Horizontal Arm Extension | 213.4 cm / 84 in |
| Weight | Shipping (with battery) 559 kg (1230 lb); unit weight without batteries 468 kg (980 lb) |
| Tube Positioning | |
| Maximum Focal Spot Height | 201.2 cm / 79.2 in |
| Minimum Focal Spot Height | 62.4 cm / 24.6 in |
| Vertical Travel | 138.4 cm / 54.5 in |
| Maximum Horizontal Extension | 107.6 cm / 42.4 in |
| Minimum Horizontal Extension | 67.0 cm / 26.4 in |
| Horizontal Travel | 40.6 cm / 16.4 in |
| Flat-Panel Digital Detector | |
| Active Image Area | 41 × 41 cm |
| Detector Dimensions | 59 cm (H) × 47 cm (W) × 2.7 cm (D) |
| Detector Weight | 5.9 kg |
| Active Matrix | 2048 × 2048 pixels |
| Image Depth | 14-bit |
| Pixel Pitch | 200 µm |
| Dynamic Range | 0.67 µR – 11 mR |
| Typical DQE | 67% @ 0 lp/mm |
| Acquisition Workstation | 15 in touchscreen monitor |
| Hard Disk | 73 GB; stores >3200 typical images |
| Acquisition Workstation | Fully processed image in <15 seconds |
| Image Display Tools | Window width and level |
| Image Display Tools | Grayscale invert |
| Image Display Tools | Interpolated zoom with roam |
| Image Display Tools | Image flips (horizontal and vertical) |
| Image Display Tools | Image rotation in 90° increments |
| Image Display Tools | Free rotation (360°) |
| Image Display Tools | Left/Right markers |
| Image Display Tools | Free-text annotation |
| Image Display Tools | Manual shuttering |
| Batteries | Shipped fully charged |
| Batteries | Sealed lead-acid, minimal-liquid “film electrolyte” design (no maintenance; won’t leak even if damaged) |
| Batteries | Nine 12 V packs in series (108 V nominal) |
| Batteries | Single battery pack powers both motor drive and X-ray system |
| Battery Charging System | Designed for charging in normal corridor or room spaces with standard ventilation |
| Battery Charging System | Not for isolated power areas (e.g., surgery) or closets |
| Minimum full charge time | 3 h 45 min |
| Battery Charging System | Charger operates in any key switch position (fast charge with system OFF, slow charge with system ON) |
| Battery Charging System | Charger status shown on message readout |
| Battery Charging System | Automatic voltage sensing prevents deep discharge and permanent battery damage |
| Battery Charging System | Automatic overcharge management to maximize battery life |
| Battery Charging System | Retractable power cord in base; hospital-grade plug for standard wall receptacles |
| Maximum line draw | 5 A @ 110 VAC (2.5 A @ 220 VAC) |
| Battery Charging System | Compatible with 110 or 220 VAC nominal, 50/60 Hz |
| Technical Control | Digital radiographic imaging system with 12.5 kW generator |
| Technical Control | Closed-loop kVp generator with microprocessor regulation for consistent kVp across battery conditions |
| Technical Control | mAs integrator measures actual mA for accuracy and linearity |
| Technical Control | Dome-type push-button switches under protective cover; technique adjustment also via touchscreen monitor |
| Technical Control | Hand switch for rotor prep, exposure, and remote field light |
| Technical Control | Removable ON/OFF key switch |
| Technical Control | Audible X-ray-on tone |
| kVp | 50–76 kVp in 2 kVp steps |
| kVp | 80–125 kVp in 5 kVp steps |
| mAs – 50–90 kVp | 0.4–320 mAs |
| mAs – 95–105 kVp | 0.4–250 mAs |
| mAs – 110–125 kVp | 0.44–200 mAs |
| Minimum Exposure Time | 4 ms |
| mA | 100 mA nominal |
| X-ray Source | |
| Tube | GE model D2651P |
| Focal Spot | 0.8 mm |
| Target Angle | 14° |
| Rotating Anode | 3 in, 3000 rpm |
| Anode Heat Storage Capacity | 140,000 HU |