The Siemens Mobilett XP is a compact, high-output portable X-ray system designed for maximum mobility, ease of positioning, and excellent image quality. Its integrated cable design keeps power and control cables enclosed within the arm, preventing tangling and simplifying cleaning—ideal for semi-sterile environments such as ICUs, CCUs, neonatal and pediatric departments, and general wards. Lightweight construction, a short wheelbase with polyurethane tires, and an effective handbrake provide smooth transportation and stable positioning, even in tight spaces and elevators. A double-articulated, fully counterbalanced X-ray arm with industry-leading horizontal extension and fully rotatable tube head allows virtually any projection with patients lying, sitting, or in between. With up to 30 kW imaging output, self-calibrating generator electronics, and exposure times as low as 1 ms, the Mobilett XP supports sharp, motion-free images for chest and general radiography applications.
Technical Specifications
| Dimensions | |
|---|---|
| Height | 61 in (155 cm) |
| Width | 23 in (57.5 cm) |
| Length | 39 in (98.7 cm) |
| Weight | ~206 kg |
| Tube Positioning | |
| Max Focal Spot Height | 2135 mm (84.1 in) |
| Min Focal Spot Height | 478 mm (18.8 in) |
| Max Horizontal Extension | 1285 mm (50.6 in) |
| Min Horizontal Extension | 282 mm (11.1 in) |
| Output | |
| Maximum | 133 kV, 225 mA |
| Maximum | 450 mA at 54 kV |
| Maximum | 30 kW at 99 kV, 303 mA, 10 ms |
| Maximum | 25 kW at 99 kV, 252 mA, 100 ms (IEC 60601-2-7) |
| Minimum mAs | 0.32 mAs |
| Line Voltage | 100–130 V ±10%, 50/60 Hz |
| Line Voltage – Impulse Current | 11 A |
| Line Voltage – Continuous Current | 0.6 A |
| Line Voltage | 200–240 V ±10%, 50/60 Hz |
| Line Voltage – Impulse Current | 19 A |
| Line Voltage – Continuous Current | 0.5 A |
| Apparent Supply Line Resistance | |
| 100–130 V | <0.6 Ω |
| 200–240 V | <2.5 Ω |
| Environment – Operation | |
| Ambient Temperature | +10 to +40°C |
| Relative Humidity | 30–75% |
| Air Pressure | 800 hPa (typical hospital range) |
| Environment – Transport/Storage | |
| Ambient Temperature | –20 to +70°C |
| Relative Humidity | 10–100%* |
| Air Pressure | up to 1060 hPa |
| Stabilization | Microprocessor-controlled adjustment and regulation during exposure |
| X-ray Tube | Siemens P135/30 R with nominal focal spot 0.8 (per IEC 60336) |
| Anode Rotation | 9000 rpm, 142 Hz |
| Anode Angle | 15° |
| Heat Storage Capacity | |
| Anode | 90,000 J (122,000 HU) |
| Tube Housing | 800,000 J (1,100,000 HU) |
| Inherent Filtration | |
| Collimator | 2.0 mm Al (at 70 kVp, min) |
| Tube Housing Assembly | 1.9 mm Al (at 70 kVp, min) |
| Total | ≥3.9 mm Al (IEC 60522) |
| Collimator | Manually adjustable multileaf collimator, rotatable ±90° |
| Collimator Light Field Intensity | Halogen lamp, ≥180 lux at 1 m SID (without DAP equipment) |
| Leakage Radiation (Example Spec) | Measured at 133 kVp / 10 mAs, 200 exposures/hour (per IEC requirements) |