The Dräger Apollo Anesthesia Workstation delivers advanced ventilation technology and ergonomic efficiency for modern operating environments. Engineered with an E-Vent Plus servo-controlled piston ventilator, the Apollo provides ICU-level precision and eliminates the need for drive gas—achieving exceptional control and quiet operation. Its low-flow anesthesia capability significantly reduces anesthetic agent consumption and overall operating costs, allowing fresh gas flow rates as low as 1 L/min without compromising patient safety. Designed with workflow optimization in mind, the Apollo integrates the breathing system, power supply, and monitoring modules into a single, compact platform. This reduces external hoses and cables, minimizing clutter and the risk of accidental disconnection during critical procedures. The user interface is intuitive, featuring a 12.1-inch full-color touchscreen, streamlined startup, and a fully automatic self-test sequence that completes in approximately three minutes. The system’s robust construction, quiet operation, and comprehensive ventilation modes—including volume, pressure, SIMV, and pressure support—make the Dräger Apollo a premier choice for hospitals seeking reliability, accuracy, and comfort for both patients and clinicians.
Technical Specifications
| Dimensions | |
|---|---|
| Height | 59 in (150 cm) |
| Width | 33.5 in (85 cm) |
| Depth | 31.5 in (80 cm) |
| Weight | 365 lb (165 kg) without vaporizers or gas cylinders |
| Power | |
| Power Consumption | 200 W |
| Operating Voltage | 100–127 VAC (–15 % / +10 %), 45–65 Hz |
| Integrated Power Backup | Minimum 30 minutes, typically 90 minutes (dependent on ventilation parameters) |
| Anesthesia Gas Supply | |
| Fresh Gas Flow | 0–12 L/min per gas (O₂, N₂O, Air) |
| O₂ Flush Flow | > 35 L/min |
| External Fresh Gas Outlet | Optional |
| Vaporizers | Dräger AutoExclusion (2 or 3 vaporizers), Dräger Interlock (2 vaporizers), or Selectatec (2 vaporizers) |
| Ventilator System | |
| Type | E-Vent Plus — electrically operated, electronically controlled piston ventilator |
| Operating Modes | Manual, Spontaneous, Volume (IPPV), Pressure (PCV), SIMV (Volume and Pressure), Pressure Support (PS), Volume Auto Flow (optional) |
| Tidal Volume Range | 20–1400 mL (5–1400 mL with advanced option) |
| Breathing Frequency | 3–100 breaths per minute |
| Inspiratory to Expiratory Ratio (I:E) | max 5:1 |
| Inspiratory Flow | up to 150 L/min |
| PEEP Range | 0–20 cmH₂O |
| System Leak Tightness | < 150 mL/min at 30 cmH₂O |
| Pressure Limits | PMAX and PINSP up to 70 cmH₂O Sensitive ORC function ensures minimum 21 vol.% O₂ or 200 mL/min with N₂O mixtures |
| Monitoring | |
| Parameters | Minute Volume (MV), Tidal Volume (VT), PEEP, Peak and Plateau Pressure, Mean Airway Pressure, Compliance (CPAT) |
| Waveforms | Pressure, Flow, and Volume; trend and numerical displays; auto alarm limit adjustment |
| Gas Monitoring | O₂, N₂O, CO₂, and volatile agents (Halothane, Enflurane, Isoflurane, Sevoflurane, Desflurane) with real-time graphing |
| Display & Controls | |
| Detail | 12.1 in full-color TFT touchscreen User-friendly layout with intuitive menus and quick-access ventilation controls |
| Breathing System | |
| Detail | Heated breathing system with integrated design to reduce hoses and connections |
| Absorber Volume | 1.5 L (reusable), 1.3 L (CLIC 800+), 1.2 L (CLIC Drägersorb Free) |
| Gas Supply Connections | |
| Pipeline Inputs | O₂, N₂O, Air |
| Cylinder Inputs | O₂, N₂O, Air Auxiliary O₂ and pressure gauge integrated at front panel |
| Interfaces & Connectivity | |
| Detail | 2 × RS-232 ports |
| Protocols | Medibus and Medibus X |
| Data Export | Pressures, Volumes, Gas Concentrations, Ventilation Settings, and Alarm Logs |
| Optional Accessories | |
| Detail | Anesthetic Gas Scavenging System (AGSS), endotracheal suction unit, writing plate, and auxiliary shelf options |