Dräger Fabius MRI

The Dräger Fabius MRI anesthesia workstation is specifically engineered for safe, reliable operation in magnetic resonance imaging (MRI) environments, supporting systems up to 1.5 and 3 Tesla.

Single Product Request Form

The Dräger Fabius MRI anesthesia workstation is specifically engineered for safe, reliable operation in magnetic resonance imaging (MRI) environments, supporting systems up to 1.5 and 3 Tesla. Combining precision performance with MRI compatibility, the Fabius MRI provides comprehensive ventilation capabilities — including volume control, pressure control, pressure support, SIMV/PS, manual, and spontaneous modes — making it suitable for a wide range of patient needs. Powered by Dräger’s E-vent piston ventilator, the Fabius MRI delivers consistent, high peak flow performance without the need for drive gas, ensuring efficient, ICU-level ventilation. A high-visibility color display with integrated warning and alarm LEDs ensures critical information is easy to read, even from a distance within the MRI suite. Designed with non-magnetic components and advanced electronic control, the system integrates all core anesthesia functions into a mobile trolley unit. It includes a 2-position vaporizer mount, pressure gauges, Dräger CLIC absorber, and Anesthetic Gas Scavenging System (AGSS) — all optimized for MRI use. The unit’s central brake, three large drawers, and compact breathing system offer ergonomic efficiency and safe mobility in confined MRI environments.

Technical Specifications

Base Unit
DetailTrolley (cart) with COSY Breathing System
Dimensions (W × H × D)39 × 55 × 35.5 in (99 × 140 × 90 cm)
Weight (without vaporizers or cylinders)365 lb (165.8 kg)
Power and Battery Backup
Input Power100–240 VAC, 50/60 Hz, 70 VA
Integrated Power OutletsIncluded
Battery Operation> 45 minutes with fully charged batteries
Ventilator System
TypeE-Vent piston ventilator, electronically controlled and electrically driven
Ventilation ModesVolume Control, Pressure Control, Pressure Support, SIMV/PS, Manual, and Spontaneous Breathing
Breathing Frequency4 – 60 bpm
PEEP0 – 20 cmH₂O
Inspiration/Expiration Ratio (I:E)4:1 to 1:4
Pressure Limiting (Pmax)15 – 70 cmH₂O
Tidal Volume (Vt)20 – 1400 mL (Volume Control)
Tidal Volume (Vt)20 – 1100 mL (SIMV/PS)
Inspiratory Pause (Tip:Ti)0 – 50 %
Inspiratory Time (SIMV)0.3 – 4.0 s
Inspiratory Pressure (Pinsp)PEEP + 5 to 65 cmH₂O
Inspiratory Flow10 – 75 L/min (Volume & Pressure Control)
Inspiratory Flow10 – 85 L/min (Pressure Support / SIMV-PS)
Pressure Support Level (ΔPPS)PEEP + 3 to 20 cmH₂O
Apnea Ventilation Frequency3 – 20 bpm or Off
Anesthesia Gas Supply
Fresh Gas Flow Indicators0.0 – 12.0 L/min
Total Fresh Gas Flowmeter0 – 10 L/min
O₂ FlushMax 75 L/min @ 87 psi (6 bar)
O₂ FlushMin 25 L/min @ 41 psi (2.8 bar)
Vaporizer MountDual-position Dräger Interlock 2 System
Compatible VaporizersDräger Isoflurane, Sevoflurane, Halothane, and Enflurane Vapors 2000
Design & ErgonomicsHigh-visibility color display with integrated alarm LEDs
Design & ErgonomicsNon-magnetic construction for MRI safety up to 3 Tesla
Design & ErgonomicsCompact breathing system with Dräger CLIC absorber
Design & ErgonomicsIntegrated Anesthetic Gas Scavenging System (AGSS)
Design & Ergonomics3 large drawers (1 lockable)
Design & ErgonomicsCentral brake and maneuverable trolley design