The GE Aisys CS2 is a state-of-the-art digital anesthesia Carestation designed to integrate precision ventilation, digital gas management, and intelligent automation in a single platform. With a 15-inch movable touchscreen monitor and CARESCAPE-inspired interface, it delivers a modern, intuitive user experience for anesthesia delivery and patient monitoring. At its core, the Aisys CS2 features digitally controlled vaporizers, ventilators, and gas delivery systems, allowing millisecond-level responsiveness to patient changes. The compact Advanced Breathing System (ABS) supports low-flow anesthesia with fast gas kinetics—offering rapid wash-in and wash-out to minimize anesthetic agent waste and enhance patient safety. Engineered for integration and performance, the Aisys CS2 offers electronic agent control, automatic data recording, and real-time gas consumption calculations. It seamlessly connects with GE’s monitoring ecosystem through standard interfaces (Ethernet, RS-232, USB, and VGA), ensuring streamlined data exchange and networked OR functionality.
Key Features
- [Design & Functional Highlights] Movable 15″ touchscreen ventilator display with CARESCAPE-style interface
- [Design & Functional Highlights] Digitally controlled vaporizer, ventilator, and gas system for precision management
- [Design & Functional Highlights] Compact ABS optimized for low-flow techniques and rapid gas exchange
- [Design & Functional Highlights] Responsive digital control measured in milliseconds
- [Design & Functional Highlights] Integrated automatic record-keeping and agent consumption tracking
Technical Specifications
| Dimensions | |
|---|---|
| Height | 52.7 in (133.9 cm) |
| Width | 26.8 in (68 cm) |
| Depth | 32.3 in (82 cm) |
| Weight | 419 lb (190 kg) |
| Ventilator Performance | |
| Pressure Range at Inlet | 240–700 kPa (35–102 psig) |
| Peak Gas Flow | 120 L/min + fresh gas flow |
| Flow Valve Range | 1 – 120 L/min |
| Flow Compensation Range | 200 mL/min – 15 L/min |
| Ventilation Modes | |
| Detail | Volume Control (VCV) Pressure Control (PCV) Pressure Control – Volume Guaranteed (PCV-VG) SIMV (Volume and Pressure) PSVPro (Pressure Support with Apnea Backup) SIMV PCV-VG CPAP + PSV |
| Ventilation Parameters | |
| Tidal Volume | 20 – 1500 mL (Volume Control, PCV-VG, SIMV, SIMV PCV-VG) |
| Minute Volume | < 0.1 – 99.9 L/min |
| Pressure (Inspired) | 5 – 60 cmH₂O (increments 1 cmH₂O) |
| Maximum Pressure (Pmax) | 12 – 100 cmH₂O (increments 1 cmH₂O) |
| Pressure Support (Psupport) | Off, 2 – 40 cmH₂O |
| Rate | 4 – 100 bpm (VCV, PCV, PCV-VG) |
| Rate | 2 – 60 bpm (SIMV, PSVPro, SIMV PCV-VG) |
| Rate | 4 – 60 bpm (CPAP+PSV) |
| I:E Ratio | 2:1 to 1:8 (increments of 0.5) |
| Inspiratory Time | 0.2 – 5.0 s (increments 0.1 s) |
| Flow Trigger | 1 – 10 L/min (increments 0.5 L/min) |
| PEEP | Off, 4 – 30 cmH₂O (increments 1 cmH₂O) |
| Volume Delivery | |
| >210 mL | ±7% |
| =210 mL | ±15 mL |
| <60 mL | ±10 mL |
| Pressure Delivery | ±10% or ±3 cmH₂O |
| PEEP Delivery | ±1.5 cmH₂O |
| Pressure Monitoring | ±5% or ±2 cmH₂O |
| Patient Spirometry & Monitoring | Pressure-Volume, Flow-Volume, and Pressure-Flow loops |
| Patient Spirometry & Monitoring | Airway pressure and flow waveforms |
| Patient Spirometry & Monitoring | Adjustable alarm limits for P_peak and M_Vexp |
| Patient Spirometry & Monitoring | Machine flow transducer-based detection |
| Sensors | |
| Flow Transducer | Variable orifice flow sensor (22 mm OD / 15 mm ID) |
| Location | Inspiratory outlet and expiratory inlet |
| Oâ‚‚ Sensor | Optional galvanic fuel cell or paramagnetic (Respiratory Module option) |
| Display | |
| Screen Size | 15 in (38 cm) |
| Resolution | 1024 × 768 |
| Mounting | Movable touchscreen ventilator monitor |
| Communication & Data Ports | |
| Detail | RS-232C serial interface Ethernet (network connectivity) Datex-Ohmeda device interface solutions port USB ports VGA output |
| Anesthetic Agent Delivery | |
| Vaporizer | Aladin2 Cassette System |
| Supported Agents | Isoflurane, Desflurane, Sevoflurane, Enflurane |
| Active Positions | 1 |
| Capacity | Isoflurane/Sevoflurane/Enflurane – 220 mL total (125 mL empty indication, 95 mL residual) |
| Capacity | Desflurane – 240 mL total (140 mL empty indication, 100 mL residual) |