GE Aisys CS2

The GE Aisys CS2 is a state-of-the-art digital anesthesia Carestation designed to integrate precision ventilation, digital gas management, and intelligent automation in a single platform.

Single Product Request Form

The GE Aisys CS2 is a state-of-the-art digital anesthesia Carestation designed to integrate precision ventilation, digital gas management, and intelligent automation in a single platform. With a 15-inch movable touchscreen monitor and CARESCAPE-inspired interface, it delivers a modern, intuitive user experience for anesthesia delivery and patient monitoring. At its core, the Aisys CS2 features digitally controlled vaporizers, ventilators, and gas delivery systems, allowing millisecond-level responsiveness to patient changes. The compact Advanced Breathing System (ABS) supports low-flow anesthesia with fast gas kinetics—offering rapid wash-in and wash-out to minimize anesthetic agent waste and enhance patient safety. Engineered for integration and performance, the Aisys CS2 offers electronic agent control, automatic data recording, and real-time gas consumption calculations. It seamlessly connects with GE’s monitoring ecosystem through standard interfaces (Ethernet, RS-232, USB, and VGA), ensuring streamlined data exchange and networked OR functionality.

Key Features

  • [Design & Functional Highlights] Movable 15″ touchscreen ventilator display with CARESCAPE-style interface
  • [Design & Functional Highlights] Digitally controlled vaporizer, ventilator, and gas system for precision management
  • [Design & Functional Highlights] Compact ABS optimized for low-flow techniques and rapid gas exchange
  • [Design & Functional Highlights] Responsive digital control measured in milliseconds
  • [Design & Functional Highlights] Integrated automatic record-keeping and agent consumption tracking

Technical Specifications

Dimensions
Height52.7 in (133.9 cm)
Width26.8 in (68 cm)
Depth32.3 in (82 cm)
Weight419 lb (190 kg)
Ventilator Performance
Pressure Range at Inlet240–700 kPa (35–102 psig)
Peak Gas Flow120 L/min + fresh gas flow
Flow Valve Range1 – 120 L/min
Flow Compensation Range200 mL/min – 15 L/min
Ventilation Modes
DetailVolume Control (VCV) Pressure Control (PCV) Pressure Control – Volume Guaranteed (PCV-VG) SIMV (Volume and Pressure) PSVPro (Pressure Support with Apnea Backup) SIMV PCV-VG CPAP + PSV
Ventilation Parameters
Tidal Volume20 – 1500 mL (Volume Control, PCV-VG, SIMV, SIMV PCV-VG)
Minute Volume< 0.1 – 99.9 L/min
Pressure (Inspired)5 – 60 cmH₂O (increments 1 cmH₂O)
Maximum Pressure (Pmax)12 – 100 cmH₂O (increments 1 cmH₂O)
Pressure Support (Psupport)Off, 2 – 40 cmH₂O
Rate4 – 100 bpm (VCV, PCV, PCV-VG)
Rate2 – 60 bpm (SIMV, PSVPro, SIMV PCV-VG)
Rate4 – 60 bpm (CPAP+PSV)
I:E Ratio2:1 to 1:8 (increments of 0.5)
Inspiratory Time0.2 – 5.0 s (increments 0.1 s)
Flow Trigger1 – 10 L/min (increments 0.5 L/min)
PEEPOff, 4 – 30 cmH₂O (increments 1 cmH₂O)
Volume Delivery
>210 mL±7%
=210 mL±15 mL
<60 mL±10 mL
Pressure Delivery±10% or ±3 cmH₂O
PEEP Delivery±1.5 cmH₂O
Pressure Monitoring±5% or ±2 cmH₂O
Patient Spirometry & MonitoringPressure-Volume, Flow-Volume, and Pressure-Flow loops
Patient Spirometry & MonitoringAirway pressure and flow waveforms
Patient Spirometry & MonitoringAdjustable alarm limits for P_peak and M_Vexp
Patient Spirometry & MonitoringMachine flow transducer-based detection
Sensors
Flow TransducerVariable orifice flow sensor (22 mm OD / 15 mm ID)
LocationInspiratory outlet and expiratory inlet
Oâ‚‚ SensorOptional galvanic fuel cell or paramagnetic (Respiratory Module option)
Display
Screen Size15 in (38 cm)
Resolution1024 × 768
MountingMovable touchscreen ventilator monitor
Communication & Data Ports
DetailRS-232C serial interface Ethernet (network connectivity) Datex-Ohmeda device interface solutions port USB ports VGA output
Anesthetic Agent Delivery
VaporizerAladin2 Cassette System
Supported AgentsIsoflurane, Desflurane, Sevoflurane, Enflurane
Active Positions1
CapacityIsoflurane/Sevoflurane/Enflurane – 220 mL total (125 mL empty indication, 95 mL residual)
CapacityDesflurane – 240 mL total (140 mL empty indication, 100 mL residual)