The GE Datex-Ohmeda E-CO Module is a compact, high-precision capnography and gas analysis module engineered to provide continuous measurement of carbon dioxide (CO₂) and respiratory parameters. Built for integration with GE patient monitoring and anesthesia systems, the E-CO module ensures accurate and reliable respiratory monitoring across diverse clinical environments, including operating rooms, ICUs, and emergency departments. Using infrared absorption technology, the module delivers precise end-tidal CO₂ (EtCO₂) readings and respiratory rate measurements in real time. Its fast response time (<400 ms) enables clinicians to detect changes in ventilation instantly, supporting better decision-making and patient safety. The user-friendly interface, configurable alarms, and durable compact design make the E-CO module a dependable choice for continuous airway monitoring in both anesthesia and critical care applications.
Key Features
- [Performance & Design Features] Continuous capnography and CO₂ monitoring with real-time waveform display
- [Performance & Design Features] Rapid response time (<400 ms) ensures fast detection of ventilatory changes
- [Performance & Design Features] Seamless integration with GE patient monitoring and anesthesia systems
- [Performance & Design Features] Advanced alarm system with customizable thresholds for respiratory safety
- [Performance & Design Features] Compact and durable construction suitable for ORs, ICUs, and transport environments
- [Performance & Design Features] Low cross-sensitivity to O₂, N₂O, and anesthetic agents for improved accuracy
- [Performance & Design Features] Plug-and-play compatibility with Datex-Ohmeda monitoring platforms
- [Key Advantages] Accurate EtCO₂ and FiCO₂ measurement for real-time ventilation assessment
- [Key Advantages] Measures O₂ and N₂O concentrations for comprehensive gas analysis
- [Key Advantages] Infrared and paramagnetic technologies ensure reliability and precision
- [Key Advantages] Easy installation and maintenance with modular integration design
- [Key Advantages] Compact footprint ideal for both fixed and mobile anesthesia setups
- [Key Advantages] Meets clinical standards for patient monitoring in anesthesia and critical care
Technical Specifications
| Dimensions | |
|---|---|
| Height | 4.4 in (11.2 cm) |
| Width | 3.0 in (7.5 cm) |
| Depth | 9.0 in (22.8 cm) |
| Weight | 3.5 lb (1.6 kg) |
| Power | |
| Power Consumption | 12.6 W (nominal), 14.6 W (momentary peak) |
| Protection Against Electrical Shock | Type BF |
| CO₂ (Capnography) | |
| Measurement Range | 0 to 15 vol% (0–15 kPa / 0–113 mmHg) |
| Rise Time | < 400 ms typical |
| Accuracy | ±(0.2 vol% + 2% of reading) |
| Example | At 5.0% reading → ±0.3 vol% |
| Gas Cross Effects | < 0.2 vol% (O₂, N₂O, anesthetic agents) |
| Respiration Rate | |
| Range | 4 to 60 breaths/min |
| Detection Criteria | 1% variation in CO₂ waveform |
| O₂ Measurement (Paramagnetic Technology) | |
| Range | 0 to 100 vol% |
| Rise Time | < 400 ms typical |
| Accuracy | ±(1 vol% + 2% of reading) |
| Gas Cross Effects | < 1 vol% (anesthetic agents) |
| Gas Cross Effects | < 2 vol% (N₂O) |
| Fi–Et Difference Resolution | 0.1 vol% |
| N₂O Measurement (Infrared Technology) | |
| Range | 0 to 100% |
| Rise Time | < 400 ms typical |
| Accuracy – 0–85% | ±(2 vol% + 2% of reading) |
| Accuracy – 85–100% | ±(2 vol% + 8% of reading) |
| Gas Cross Effects | < 2 vol% (anesthetic agents) |