GE Datex-Ohmeda E-CO Module

The GE Datex-Ohmeda E-CO Module is a compact, high-precision capnography and gas analysis module engineered to provide continuous measurement of carbon dioxide (CO₂) and respiratory parameters.

Single Product Request Form

The GE Datex-Ohmeda E-CO Module is a compact, high-precision capnography and gas analysis module engineered to provide continuous measurement of carbon dioxide (CO₂) and respiratory parameters. Built for integration with GE patient monitoring and anesthesia systems, the E-CO module ensures accurate and reliable respiratory monitoring across diverse clinical environments, including operating rooms, ICUs, and emergency departments. Using infrared absorption technology, the module delivers precise end-tidal CO₂ (EtCO₂) readings and respiratory rate measurements in real time. Its fast response time (<400 ms) enables clinicians to detect changes in ventilation instantly, supporting better decision-making and patient safety. The user-friendly interface, configurable alarms, and durable compact design make the E-CO module a dependable choice for continuous airway monitoring in both anesthesia and critical care applications.

Key Features

  • [Performance & Design Features] Continuous capnography and CO₂ monitoring with real-time waveform display
  • [Performance & Design Features] Rapid response time (<400 ms) ensures fast detection of ventilatory changes
  • [Performance & Design Features] Seamless integration with GE patient monitoring and anesthesia systems
  • [Performance & Design Features] Advanced alarm system with customizable thresholds for respiratory safety
  • [Performance & Design Features] Compact and durable construction suitable for ORs, ICUs, and transport environments
  • [Performance & Design Features] Low cross-sensitivity to O₂, N₂O, and anesthetic agents for improved accuracy
  • [Performance & Design Features] Plug-and-play compatibility with Datex-Ohmeda monitoring platforms
  • [Key Advantages] Accurate EtCO₂ and FiCO₂ measurement for real-time ventilation assessment
  • [Key Advantages] Measures O₂ and N₂O concentrations for comprehensive gas analysis
  • [Key Advantages] Infrared and paramagnetic technologies ensure reliability and precision
  • [Key Advantages] Easy installation and maintenance with modular integration design
  • [Key Advantages] Compact footprint ideal for both fixed and mobile anesthesia setups
  • [Key Advantages] Meets clinical standards for patient monitoring in anesthesia and critical care

Technical Specifications

Dimensions
Height4.4 in (11.2 cm)
Width3.0 in (7.5 cm)
Depth9.0 in (22.8 cm)
Weight3.5 lb (1.6 kg)
Power
Power Consumption12.6 W (nominal), 14.6 W (momentary peak)
Protection Against Electrical ShockType BF
CO₂ (Capnography)
Measurement Range0 to 15 vol% (0–15 kPa / 0–113 mmHg)
Rise Time< 400 ms typical
Accuracy±(0.2 vol% + 2% of reading)
ExampleAt 5.0% reading → ±0.3 vol%
Gas Cross Effects< 0.2 vol% (O₂, N₂O, anesthetic agents)
Respiration Rate
Range4 to 60 breaths/min
Detection Criteria1% variation in CO₂ waveform
O₂ Measurement (Paramagnetic Technology)
Range0 to 100 vol%
Rise Time< 400 ms typical
Accuracy±(1 vol% + 2% of reading)
Gas Cross Effects< 1 vol% (anesthetic agents)
Gas Cross Effects< 2 vol% (N₂O)
Fi–Et Difference Resolution0.1 vol%
N₂O Measurement (Infrared Technology)
Range0 to 100%
Rise Time< 400 ms typical
Accuracy – 0–85%±(2 vol% + 2% of reading)
Accuracy – 85–100%±(2 vol% + 8% of reading)
Gas Cross Effects< 2 vol% (anesthetic agents)