Siemens Mobilett XP

The Siemens Mobilett XP is a compact, high-output portable X-ray system designed for maximum mobility, ease of positioning, and excellent image quality.

Single Product Request Form

The Siemens Mobilett XP is a compact, high-output portable X-ray system designed for maximum mobility, ease of positioning, and excellent image quality. Its integrated cable design keeps power and control cables enclosed within the arm, preventing tangling and simplifying cleaning—ideal for semi-sterile environments such as ICUs, CCUs, neonatal and pediatric departments, and general wards. Lightweight construction, a short wheelbase with polyurethane tires, and an effective handbrake provide smooth transportation and stable positioning, even in tight spaces and elevators. A double-articulated, fully counterbalanced X-ray arm with industry-leading horizontal extension and fully rotatable tube head allows virtually any projection with patients lying, sitting, or in between. With up to 30 kW imaging output, self-calibrating generator electronics, and exposure times as low as 1 ms, the Mobilett XP supports sharp, motion-free images for chest and general radiography applications.

Technical Specifications

Dimensions
Height61 in (155 cm)
Width23 in (57.5 cm)
Length39 in (98.7 cm)
Weight~206 kg
Tube Positioning
Max Focal Spot Height2135 mm (84.1 in)
Min Focal Spot Height478 mm (18.8 in)
Max Horizontal Extension1285 mm (50.6 in)
Min Horizontal Extension282 mm (11.1 in)
Output
Maximum133 kV, 225 mA
Maximum450 mA at 54 kV
Maximum30 kW at 99 kV, 303 mA, 10 ms
Maximum25 kW at 99 kV, 252 mA, 100 ms (IEC 60601-2-7)
Minimum mAs0.32 mAs
Line Voltage100–130 V ±10%, 50/60 Hz
Line Voltage – Impulse Current11 A
Line Voltage – Continuous Current0.6 A
Line Voltage200–240 V ±10%, 50/60 Hz
Line Voltage – Impulse Current19 A
Line Voltage – Continuous Current0.5 A
Apparent Supply Line Resistance
100–130 V<0.6 Ω
200–240 V<2.5 Ω
Environment – Operation
Ambient Temperature+10 to +40°C
Relative Humidity30–75%
Air Pressure800 hPa (typical hospital range)
Environment – Transport/Storage
Ambient Temperature–20 to +70°C
Relative Humidity10–100%*
Air Pressureup to 1060 hPa
StabilizationMicroprocessor-controlled adjustment and regulation during exposure
X-ray TubeSiemens P135/30 R with nominal focal spot 0.8 (per IEC 60336)
Anode Rotation9000 rpm, 142 Hz
Anode Angle15°
Heat Storage Capacity
Anode90,000 J (122,000 HU)
Tube Housing800,000 J (1,100,000 HU)
Inherent Filtration
Collimator2.0 mm Al (at 70 kVp, min)
Tube Housing Assembly1.9 mm Al (at 70 kVp, min)
Total≥3.9 mm Al (IEC 60522)
CollimatorManually adjustable multileaf collimator, rotatable ±90°
Collimator Light Field IntensityHalogen lamp, ≥180 lux at 1 m SID (without DAP equipment)
Leakage Radiation (Example Spec)Measured at 133 kVp / 10 mAs, 200 exposures/hour (per IEC requirements)