The Siemens Mobilett Plus HP is a high-performance portable X-ray system offering dual power-source capability for maximum flexibility. It can operate from its built-in rechargeable batteries or directly from a standard AC power source, ensuring uninterrupted imaging in busy clinical environments. Its compact, lightweight design, long-reaching swivel arm, and unique turn-plate base allow the system to pivot 360°, enabling fast and precise positioning around the patient. With high radiographic output and short exposure times, the Mobilett Plus HP helps reduce motion artifacts and improve image sharpness across a variety of applications, including ICU, coronary care, neonatal, burn units, surgery, pediatrics, and emergency departments. Digital controls on the LED console, battery charge indicators, and clear operator information displays support efficient and safe mobile radiography.
Technical Specifications
| Dimensions | |
|---|---|
| Height | 6.8 in (157 cm)* |
| Width | 22 in (65.4 cm) |
| Length | 41 in (104 cm) |
| Weight | 628 lb (285 kg) (*Height value as provided.) |
| Tube Positioning | |
| Maximum Focal Spot Height | 84.4 in (2145 mm) |
| Minimum Focal Spot Height | 18.5 in (470 mm) |
| Maximum Horizontal Extension | 50.4 in (1280 mm) |
| Minimum Horizontal Extension | 12.4 in (316 mm) |
| Mains Voltage | 100–130 V ±10%, 50/60 Hz, fuse 15 A |
| Mains Voltage | 200–240 V ±10%, 50/60 Hz, fuse 10 A |
| Batteries | 16 sealed, maintenance-free lead batteries |
| Batteries – Usable battery storage | 20,000 mAs |
| Batteries | Automatic battery charging with charge indicator |
| Batteries | Mains operation with reduced output |
| Output | 30 kW at 96 kV, 10 ms |
| Output | 20 kW at 102 kV, 100 ms (IEC 601-2-7/1987) |
| Maximum kV | 133 kV |
| Maximum mA | 450 mA |
| kVp Range & Accuracy | |
| Range | 40–133 kV in 25 steps |
| Accuracy | <±3 kV (40–60 kV) |
| Accuracy | <±5% (60–133 kV) |
| mAs Range | 0.5–360 mAs / 40–50 kV |
| mAs Range | 0.5–320 mAs / 52–73 kV |
| mAs Range | 0.5–280 mAs / 77–90 kV |
| mAs Range | 0.5–25 mAs / 96 kV |
| mAs Range | 0.5–16 mAs / 125 kV |
| mAs Range | 0.5–12.5 mAs / 133 kV |
| Exposure Times | |
| Shortest | 1 ms |
| Maximum | 4.0 s |
| Power Cord | |
| Length | 6 m automatic retractable cord |
| Tube Housing Assembly | |
| X-ray Tube | Siemens P135/30 R with nominal focal spot |
| Anode Rotation | 8500 rpm, 150 Hz; 15° anode angle |
| Anode Heat Storage Capacity | 90,000 J (122,000 HU) |
| Tube Housing Capacity | 800,000 J (1,100,000 HU) |
| Inherent Filtration | min. 3.8 mm Al (IEC 522/1976) |
| Collimator | |
| Type | Manually adjustable multi-leaf, rotatable ±90° |
| Light Field Intensity | Halogen lamp, min. 180 lux at 1 m SID |
| Tube Housing Thermal Data | |
| Maximum Temperature | 60°C |
| Detail | Max 2000 mAs/h continuous rating; nominal 75 W until covers reach 60°C |
| Motor Drive | |
| Drive Speed | Low/high, 0.5–1 m/s (adjustable) |
| Maximum Incline | 7° |