The GE Optima XR220amx is a self-contained, battery-operated mobile digital X-ray system designed for point-of-care radiographic exams when transporting patients is not safe or practical. It is well-suited for routine radiographic procedures in radiology departments, ICUs, CCUs, emergency rooms, operating rooms, orthopedic areas, pediatrics, patient rooms, and clinics. Equipped with GE’s FlashPad lightweight detector, the XR220amx enables rapid, high-resolution digital imaging with minimal setup. The system’s “stand-by” mode supports near-instant readiness, while its robust battery system and optional 30 kW generator provide power and flexibility for a wide range of clinical procedures, including bariatric applications. Automatic charging algorithms and bin-based detector charging streamline workflow and help extend component life, while the unit’s integrated status displays and power architecture maintain dependable operation throughout the day.
Key Features
- Battery-operated mobile digital X-ray system
- FlashPad detector (<4.5 kg / 10 lb including battery)
- “Stand-by” mode allows exposure within 25 seconds
- 24/7 availability: can image on battery or while charging
- Detector supports up to 160 kg (352 lb) distributed load (bariatric-capable)
- Detector battery charges automatically when detector is docked in the bin
- Automatic charging algorithms support partial-charge cycling
- 15 kW (nominal) generator with optional 30 kW (nominal) upgrade
Technical Specifications
| Wireless Digital Detector | |
|---|---|
| Battery capacity | up to 50 exposures/hour; approx. 3 hours of use per charge |
| Detector Type | Single-panel (non-tiled) amorphous silicon with CsI scintillator |
| Image Area | 40.4 × 40.4 cm (15.9 × 15.9 in) |
| Active Matrix | 2024 × 2024 pixels |
| Pixel Pitch | 200 µm |
| Typical Upper Dynamic Range | 7.8 mR |
| Typical DQE | 68% @ 0 lp/mm |
| Detail | Two integrated handgrips |
| Detector Environmental Conditions | |
| Operating Temperature | 15–35°C |
| Non-Operating Temperature | –0 to 50°C (max change 20°C/hour) |
| Operating Humidity | 10–80% RH, non-condensing (max change 30%/hour) |
| Non-Operating Humidity | 10–95% RH, non-condensing (max change 30%/hour) |
| Power | |
| Input | 100–240 V nominal, 50/60 Hz |
| System Battery Status Display | Integrated |
| X-ray Source | |
| Tube Voltage (Radiographic) | 40–150 kV |
| Focal Spot Size (IEC 60336) – Large Focus | 1.2 mm |
| Focal Spot Size (IEC 60336) – Small Focus | 0.6 mm |
| Anode Rotation Speed (minimum) | 3200 min⁻¹ |
| Permanent Filtration | 0.9 mm Al @ 75 kV (IEC 60522:1999) |
| Maximum Tube Current – Large Focus | 500 mA |
| Maximum Tube Current – Small Focus | 200 mA |
| Maximum Continuous Heat Dissipation | 170 W (238 HU/s) without air circulator |
| Maximum Anode Heat | 140 kHU |
| Exposure Time Range | 4.0 ms – 6.0 s |