GE AMX 4 Plus

The GE AMX 4 Plus is an advanced, battery-operated portable X-ray system designed for consistent, full-scale radiographic image quality at the point of care.

Single Product Request Form

The GE AMX 4 Plus is an advanced, battery-operated portable X-ray system designed for consistent, full-scale radiographic image quality at the point of care. Its compact, easy-to-maneuver design allows technologists to obtain diagnostic images in ICUs, CCUs, emergency departments, ORs, pediatric and neonatal units, orthopedic wards, and clinics—without the need to plug in before making exposures. Dual drive motors and oversized casters provide smooth maneuverability, while microprocessor-based operator controls help ensure accurate technique selection and system status feedback. The AMX 4 Plus supports up to 50 high-quality exposures on a single charge and offers a robust manual collimator with built-in SID controls. This system is engineered for reliability, productivity, and repeatable image quality in even the most demanding mobile imaging environments.

Key Features

  • Analog portable X-ray system in a compact, maneuverable package
  • Up to 50 high-quality exposures per full charge
  • Dual-drive motors and oversized casters for easy movement
  • Microprocessor-based operator controls and message display
  • Ergonomic hand switch controls for prep and exposure
  • Compact manual collimator with built-in SID button on collimator face

Technical Specifications

General
Height70 in (177.8 cm) for models 2169360-6, 2236420-6, 2275938-6, -12, -13, -14, -15
Height76 in (193 cm) for all other models
Width25-3/16 in (64 cm)
Length45-3/8 in (115.3 cm)
Weight1080 lb (490 kg)
Environmental Limits
Operating Temperature59–100°F (15–38°C), up to 80% non-condensing humidity
Storage Temperature–40 to +140°F (–40 to +60°C)
Maximum Operating Altitude8000 ft (2440 m)
Movements (tube vertical movement measured at focal spot, arm extended)
Vertical Range≥46.5 in
Lowest Position26.1 in
Highest Position72.6 in
Horizontal movement at focal spot relative to column face24 in (61.0 cm) min to 40 in (101.6 cm) max
Tube Column Rotation (from horizontal arm latch)±270°
Tube and Yoke Rotation around Horizontal Arm (from tube port down)360° range; detents at ±90° and ±180°
Tube Trunnion Rotation (from tube port down)120° range; forward 110°, backward 10°; detents at 0° and 90°
Collimator Rotation (from front with tube port down)180° range; right 90°, left 90°; detents at 0° and 90°
Drive Speed
Drive Speed264 ft (670.5 cm) per min ±25%
Maneuvering Speed30–60% of drive speed
kVp Accuracy
kVp rise time (10% to 90% max kVp)≤1.2 ms
kVp fall time (90% max kVp to 20 kV)≤2.5 ms
Accuracy to selected kVp±8% for first 20 ms, ±5% after 20 ms (within bar graph battery indicator range)
Collimator Function
TypeManual
DetailMinimum source-to-skin distance >30 cm via skin spacers at beam exit
DetailFull 17 × 17 in (43 × 43 cm) coverage at 40 in (1.02 m) SID
Minimum inherent filtration2.0 mm Al equivalent at 100 kVp
Collimator Light Field IntensityAverage illumination at 100 cm (39.37 in) from focal spot ≥16 foot-candles (≈170 lux)
Generator Performance & Tube
Focal Spot0.75 mm (NEMA)
TubeGE X-ray tube model HRT09, 3 in rotating anode
Anode Heat Capacity275,000 HU
Target Angle15°
Anode Speed3000 rpm (low speed)
DetailClosed-loop kVp generator design with microprocessor regulation for constant, accurate kVp across battery conditions
Battery SpecificationsNine 12.9 V batteries connected in series (≈116 V at full charge)
Battery Capacity (Measured with fully charged, healthy batteries)
Method 1 – Exams>20 typical exams, where one exam =
Method 1 – ExamsTwo 70 kVp, 10 mAs exposures
Method 1 – Exams7 s prep (rotor/filament)
Method 1 – Exams25 s field light
Method 1 – Exams5 min driving
Method 1 – Exams9 min idle
Method 2 – X-ray Exposures≥165 exposures at 100 kVp, 100 mAs (each with 4 s prep and 30 s idle)
Method 3 – Drive Time~140 minutes of continuous drive time (may decrease on carpet/ramps)
Method 4 – Idle Time~23.3 hours of continuous idle (ON, no use)
Method 5 – Varied Usage (Capacity Formula){(idle time in minutes) × 3} + {(drive time in minutes) × 30} + {(field light time in minutes) × 25} + {(prep time in minutes) × 30} + {(exposure energy*) × 2.17} = 4200 *exposure energy = cumulative {(kVp × mAs) / 1000}
Typical pediatric exam withTwo 70 kVp, 0.8 mAs exposures
Typical pediatric exam with3 s prep
Typical pediatric exam with15 s field light
Typical pediatric exam with1 min drive
Typical pediatric exam with5 min idle
Typical pediatric exam withEach exam uses ≈60.7 “capacity units,” yielding ≈69 exams from a full charge (4200 / 60.7).